Test & measurement · Infrared sensing+886 2 8797 1799

VIGO PHOTONICS · SMART TAIWAN

PVA-3TE-5 AF anti-fringing detector for laser spectroscopy

PVA-3TE-5-AF1x1AR-TO8-wAl2O3AR-70

For laser gas spectroscopy affected by etalon / fringing baseline variations, evaluate the anti-fringing design of PVA-3TE-5 AF. Confirm the laser wavelength and full part number before choosing external amplification, cooling and optics.

PVA-3TE-5-AF1x1AR-TO8-wAl2O3AR-70

Specifications and configuration

Key specifications
Material / detectionInAsSb / photovoltaic
Typical spectral range1.64–5.8 μm (10% response)
Anti-fringing design band1.4–4.0 μm
Peak wavelength4.0 μm (typ.)
Active area1 × 1 mm; no optical immersion
Cooling / chip temperature3TE / approximately 210 K
Package / acceptance angle2° wedged TO8 cap / approximately 70°
Typical D*1.4 × 10¹⁰ cm·Hz½/W (λpeak, 20 kHz)
Typical current responsivity2.0 A/W (λpeak)
Typical time constant124 ns

Tamb = 293 K; Tchip = 210 K; Vb = 0 V.

Typical values are not guaranteed limits. Refer to the original English datasheet for full test conditions, minimum / maximum values and operating limits. D* figures measured at different frequencies are not directly comparable.

Manufacturer spectral response

PVA-3TE-5-AF1x1AR-TO8-wAl2O3AR-70 — manufacturer typical spectral response; conditions shown in the graph and datasheet
VIGO Photonics typical spectral curve; evaluate at the stated conditions and intended operating wavelength.

Configuration checks before evaluation

  • Check the anti-fringing design band separately from the full spectral response.
  • This TO8 detector needs external electronics; review the listed AIP, PIP, MIP or SIP-TO8 amplifier families for your requirements.
  • Confirm TEC control, heat dissipation, spot and incidence angle; detector time constant is not complete-receiver bandwidth.

Original English documents and resources

Application guidance checked on 18 November 2026 against the manufacturer product page and linked datasheet. Candidate applications remain subject to the complete optical and readout design. Datasheet: V1.0, 13 November 2025, Preliminary.

Start with your measurement requirements

Share the model or application, wavelength, signal bandwidth, quantity and desired schedule. Unknown items can be discussed during selection.

Submit a VIGO inquiry